Lead apron attenuation: Bushong lists 75% to 88% for 0.5 mm Pb at 100 and 75 kVp, but other sources run higher
Key takeaways
- Thicker lead protects more. Higher kVp lowers protection.
- Bushong's table and direct-beam measurements give different percentages, so learn the table your program uses.
- 21 CFR 1020.32 sets no apron minimum. The Health Physics Society says state regulations do.
- Bushong and Sherer give 0.25 mm Pb as the minimum. NCRP Report 102 and Sherer give 0.5 mm Pb for fluoroscopy.
There isn’t one correct percentage. Good sources disagree, so learn the direction first: more lead means more protection, and higher kVp means less.
What does the textbook table say?
Bushong’s table, repeated in Sherer:
| Lead equivalent | 50 kVp | 75 kVp | 100 kVp | Weight |
|---|---|---|---|---|
| 0.25 mm Pb | 97% | 66% | 51% | 1 to 5 kg |
| 0.50 mm Pb | 99.9% | 88% | 75% | 3 to 7 kg |
| 1.00 mm Pb | 99.9% | 99% | 94% | 5 to 12 kg |
Why do other sources say about 97%?
Other sources report much higher figures than that table:
- Livingstone and Varghese (2018) measured a mean of 97% for 0.5 mm aprons and 90% for 0.25 mm aprons, at 100 kVp.
- Sherer itself reports about 3.2% transmission through a 0.5 mm apron at 100 kVp, or about 97% attenuation. That sits in the same book as the 75% table.
- Bontrager says a 0.5 mm apron reduces exposure by a factor of 50 over the diagnostic x-ray range. That works out to about 98%, with no kVp or beam stated.
The 2018 study and Sherer’s figure are of a direct beam. Neither is a scatter figure, and aprons mostly face scatter. Bushong’s table states no beam condition, so it may not be measuring the same thing. We can’t tell you which number an exam item would use. ARRT® says it avoids content where resources conflict, and its outline lists “attenuation properties” without numbers. If your program teaches the table, know the table.
An earlier version of this page told you 95% to 99% was a myth for a 0.5 mm apron. The measurements don’t support saying that.
Is there a federal minimum?
None that we found in 21 CFR Part 1020 or 10 CFR Part 20. An earlier version of this page claimed 21 CFR 1020.32 set 0.5 mm. That section is about equipment and never mentions aprons. See our thyroid shield page for what it does cover.
The Health Physics Society says minimums are set in state regulations, commonly 0.25 or 0.5 mm. We didn’t open any state regulation. ARRT’s outline points to NCRP Report 102 for minimum lead equivalents. We read a public copy: it says protective aprons of at least 0.5 mm lead equivalent shall be worn in the fluoroscopy room by each person except the patient. Our text search found no 0.25 mm apron minimum there, so that figure comes from textbooks. By source:
- Apron, minimum: 0.25 mm Pb (Bushong, Sherer).
- Apron, fluoroscopy and interventional: 0.5 mm Pb (NCRP Report 102 for fluoroscopy, Sherer).
- Gloves: at least 0.25 mm Pb when hands are near the field (Sherer). NCRP Report 102 says a fluoroscopist’s hand shall not be placed in the useful beam unless the beam is attenuated by the patient and a protective glove of at least 0.25 mm lead equivalent.
These match our Radiation Protection guide. Many facilities stock only 0.5 mm so nobody picks the wrong one. Sherer calls it a good balance of protection and weight.
What else do you need to know?
- Wraparound aprons make sense if your back may face the source.
- Storage: hang aprons on a rack. Folding them can crack the lead.
- Checks: Bushong says to fluoroscope aprons and gloves for cracks at least once a year, and that apparel with cracks, tears or holes may require replacement. Our advice: stop using a cracked one and report it.
The physics is in the Radiation Physics and Radiobiology chapter. Drill the rest with the Radiation Protection practice set.
Frequently asked questions
How much radiation does a 0.5 mm lead apron block?
Bushong's table says 88% at 75 kVp and 75% at 100 kVp. A 2018 study measured about 97% at 100 kVp in the primary beam, and Sherer reports about 3.2% transmission at 100 kVp. Bontrager says a factor of 50 over the diagnostic range, which is about 98%. The sources conflict, and none of these gives a figure for scatter.
Is there a federal minimum thickness for lead aprons?
Not in 21 CFR 1020.32, which doesn't mention aprons. The Health Physics Society says minimums are set in state regulations, commonly 0.25 or 0.5 mm.
Is a 0.25 mm lead apron enough?
Bushong and Sherer give 0.25 mm Pb as the minimum for an apron. Sherer says fluoroscopy and interventional work need 0.5 mm Pb, NCRP Report 102 sets 0.5 mm in the fluoroscopy room, and many facilities stock only 0.5 mm.
How often are lead aprons checked?
Bushong says aprons and gloves should be fluoroscoped for cracks at least once a year.
Sources
Found an error? Email [email protected] with the page and the source.
- NCRP Report 102, Medical X-Ray, Electron Beam and Gamma-Ray Protection for Energies Up to 50 MeV, 1989, Sections 3.3.4(p) and (q) (public copy at law.resource.org) Official
- Livingstone and Varghese, A simple quality control tool for assessing integrity of lead equivalent aprons, Indian J Radiol Imaging, 2018 Journal
- 21 CFR 1020.32, Fluoroscopic equipment Regulation
- Health Physics Society, Ask the Experts, question 14090 Society