Safety

ARRT Radiation Protection: ALARA, Dose Limits and Shielding

Radiation Protection on the ARRT® Radiography exam splits into two parts: keeping patient dose down, and keeping your own dose down. The registry tests both, and it uses SI units as its primary units. This guide says which numbers come from a federal rule, which come from NCRP by way of the textbooks listed under sources, and where the books disagree.

This page is a study reference, not clinical guidance. Follow your state rules, your facility’s radiation safety program and your training.

How many Radiation Protection questions are on the ARRT exam?

Radiation Protection has 29 of 200 scored questions through February 28, 2027, and 25 from March 1, 2027. That is a drop from 14.5% to 12.5%. It sits in the Safety section, which falls from 50 to 46. The other Safety category, Radiation Physics and Radiobiology, stays at 21.

ARRT doesn’t publish how many questions cover each topic, so we won’t guess at a split. Not sure which section is weakest? Try the free practice test. For the whole exam plan, read how to pass the ARRT Radiography exam.

Does ALARA apply to patients or only to personnel?

ALARA means as low as reasonably achievable. In ARRT’s outlines, it applies to personnel. Both the 2022 and 2027 outlines list “Personnel Protection (ALARA)” and carry a note dated August 24, 2016. It says “the ALARA concept is specific to personnel protection and is listed only for that section.”

The same note says radiographers must still apply protection principles to minimize bioeffects for both patients and personnel. Patient dose is simply filed under a different heading, “Minimizing Patient Exposure”.

You will see the term used more widely elsewhere:

  • The NRC’s rule, 10 CFR 20.1101, uses ALARA for occupational doses and doses to members of the public.
  • AAPM’s shielding statement describes patient shielding as long considered “consistent with the ALARA principle”.
  • Ehrlich and Coakes’s patient care textbook applies ALARA to both patients and workers.

So on the exam, tie the word ALARA to your own protection, and treat patient dose reduction as its own topic. The right action rarely changes. Only the label does. That is our reading of the note, not an ARRT statement about how a question is scored.

What are time, distance and shielding?

These are the three basic methods of personnel protection in the outline.

Time. Less beam-on time means less dose. In fluoroscopy, that means short pedal taps, pulsed mode and last-image hold.

Distance. Intensity falls with the square of the distance from the source. Double your distance and intensity drops to one-fourth. Going from 2 m to 4 m cuts it to one-fourth, and going to 6 m cuts it to one-ninth. Sherer puts scatter at 1 m, at 90 degrees to the beam, at about 1/1000 of the primary beam.

An x-ray beam spreading from the focal spot: 4 mGya at 90 cm and 1 mGya at 180 cm, over four times the area
Double the distance, one-fourth the intensity. The patient isn't a point source, so scatter in a real room won't follow the formula exactly.

Shielding. Put attenuating material between you and the source. The 2027 outline adds the words “need and appropriate use” here. Lead is the usual material, and barriers count too.

Your main source of dose is scatter from the patient, so the patient is a source in the outline’s own list. Bushong says radiology staff should never hold patients. Use a mechanical device first. If someone must hold, Sherer and Bontrager say to use a person who isn’t a radiation worker, in protective apparel and never pregnant. Bontrager also rules out anyone under 18.

What are the dose limits for radiographers?

These are the numbers and where each one comes from. “NRC rule” means we read it in 10 CFR Part 20. “NCRP” means Sherer and Bushong report it from NCRP Report 116. We read those books, not the report itself.

NRC occupational and public dose limits
Who or whatLimit
Occupational, whole body (effective dose)50 mSv a year
Occupational, skin or extremity500 mSv a year
Occupational, lens of the eye (NRC rule)150 mSv a year
Embryo-fetus, declared pregnancy5 mSv for the entire pregnancy
Member of the public1 mSv a year

NRC limits from 10 CFR 20, not ARRT limits. NCRP recommends a lower lens limit. See the lens section.

Who or whatLimitBasis
Occupational, whole body50 mSv (5 rem) a year, effective doseNRC rule, 20.1201
Occupational, skin or extremity500 mSv (50 rem) a yearNRC rule, 20.1201
Occupational, lens of the eye150 mSv (15 rem) a yearNRC rule, 20.1201. See the next section.
Member of the public1 mSv (0.1 rem) a yearNRC rule, 20.1301
Embryo-fetus, declared pregnancy5 mSv (0.5 rem) for the entire pregnancyNRC rule, 20.1208
Embryo-fetus, per month0.5 mSv (0.05 rem)NCRP. The NRC rule instead says to avoid substantial variation above a uniform monthly rate.
Occupational, cumulative10 mSv × age in yearsNCRP. Not in the NRC rule.
Under 18, occupational5 mSv (0.5 rem) a year, 10% of the adult limitNRC rule, 20.1207
Under 18, education and training1 mSv (0.1 rem) a yearNCRP

Four things to remember:

  • The NRC writes Part 20 for its licensees. State and facility rules decide what binds you at work.
  • The ARRT outline cites NCRP for these topics, and NCRP limits and NRC limits aren’t always the same. Bushong and Sherer treat 0.5 mSv a month as a limit, and the NRC rule doesn’t.
  • These limits don’t apply to patients. The public limit in 20.1301 excludes doses from your own medical care.
  • Watch for one book conflict. Ehrlich and Coakes prints 50 mSv for the pregnant worker over 9 months. The NRC rule, Sherer, Bushong and Bontrager all say 5 mSv, so use 5 mSv.

What is the dose limit for the lens of the eye?

This is the one number that is moving, so here is exactly what we could and couldn’t check.

Confirmed:

  • The NRC limit is 150 mSv (15 rem) a year, in 10 CFR 20.1201(a)(2)(i).
  • NCRP Commentary 26 (2016) recommended lowering NCRP’s limit from an equivalent dose of 150 mSv to an absorbed dose of 50 mGy a year. We read this in the abstract of the published overview, not in the commentary itself. Note the unit: mGy, not mSv.
  • ICRP recommends 20 mSv a year averaged over 5 years, with no single year above 50 mSv.
  • The 2022 ARRT outline cites NCRP Report 116 for personnel monitoring. The 2027 outline cites NCRP Report 180, plus 21 CFR. We haven’t read that report and give no value from it.
  • Sherer (2018) and Bushong (2017) both print 150 mSv from NCRP Report 116. Neither mentions Commentary 26 or Report 180, so don’t treat them as current on this one number.

Not confirmed:

  • Which value ARRT’s items use. ARRT doesn’t publish that.

Our advice, marked as judgment: through February 28, 2027, know 150 mSv. For an exam on or after March 1, 2027, know both numbers and read the stem. If it names the NRC or 10 CFR, the answer is 150 mSv. If it asks for the current NCRP recommendation, the latest value we could confirm is 50 mGy. Ask your program director what your program teaches.

Either way, protect your eyes. The NRC paper notes that lens dose can be well above effective dose when an apron shields the trunk, and that leaded glasses should have side shields.

What lead equivalent do protective devices need?

ARRT’s outline cites NCRP Report 102 for minimum lead equivalents. We read a public copy: it says at least 0.5 mm for aprons worn in the fluoroscopy room. It also says a fluoroscopist’s hand shall not be placed in the useful beam unless the beam is attenuated by the patient and a protective glove of at least 0.25 mm, so a glove alone doesn’t make it acceptable. The table’s values come from Sherer. Bushong agrees on the apron minimum, the curtain and the slot cover.

DeviceLead equivalent
Apron, minimum0.25 mm Pb
Apron, fluoroscopy and interventional0.5 mm Pb
Thyroid shield0.5 mm Pb
Lead glasses0.35 mm Pb
Lead gloves0.25 mm Pb
Fluoroscopy curtain or sliding panelAt least 0.25 mm Pb
Bucky slot coverAt least 0.25 mm Pb

On fluoroscopy, Sherer calls 0.5 mm the minimum, NCRP Report 102 also says at least 0.5 mm, and Bushong says at least 0.25 mm is required and 0.5 mm is normal. 21 CFR 1020 doesn’t mention aprons. The Health Physics Society says state regulations set the minimums, commonly 0.25 mm in some states and 0.5 mm in others. We didn’t open any state regulation, so ask your radiation safety officer what applies. See lead apron attenuation for the percentages, and whether a thyroid shield is required for what the rules say.

Check aprons and gloves for cracks at least once a year, by fluoroscopy or by high-kVp radiography. Bushong and Sherer both give this interval. Sherer calls it a regulatory requirement but names no regulation, and we didn’t find one. Hang aprons on a rack, since folding can crack them. If a check finds cracks, tears or holes, Bushong says the apparel may require replacement.

Where do you wear a dosimeter?

Types named in the 2027 outline are TLD, OSL and DIS. A dosimeter measures dose. It doesn’t protect you.

Figure wearing a lead apron and thyroid shield with a primary dosimeter at the collar outside the apron, a second at the waist under the apron, and a ring on the hand
Collar outside the apron, second badge at the waist under it. The 0.5 mSv monthly figure is NCRP's, not the NRC rule, as noted above.
  • With an apron: at the collar, outside the apron. All the books we read agree. Under the apron, it would read falsely low and hide the dose to your head and eyes.
  • Without an apron, in routine radiography: the books disagree. Sherer and Ehrlich and Coakes say collar level. Bushong and Bontrager accept the front of the body at waist or chest level, as our chapter does. Your program or radiation safety officer decides at work. Our judgment: the collar works in both settings, and Sherer stresses wearing the badge in the same place every time.
  • Second badge in long interventional work: some facilities add one at the waist, under a wraparound apron. See the two-badge formula.
  • Declared pregnancy: a second dosimeter at the waist, under the apron, with its record kept separately.
  • Who must be monitored: the NRC rule requires it for adults likely to get more than 10% of the limits, and for declared pregnant workers likely to get more than 1 mSv (0.1 rem) during the pregnancy.

Declaring a pregnancy is voluntary and must be in writing. The NRC limit then protects the embryo-fetus, and the worker’s own limit is unchanged. We found no rule that bars pregnant staff from the room. Sherer and Bontrager say a pregnant radiographer can keep working in fluoroscopy under normal safety practice, so the facility’s program decides any work changes.

How do you reduce dose in fluoroscopy?

Bushong says at least 95% of a radiographer’s occupational dose comes from fluoroscopy and mobile radiography. For patients, the picture differs: CT was about 62% of collective patient dose in the U.S. in 2016 (444,000 of 717,000 person-Sv, from NCRP Report 184 data).

These are the fluoroscopy facts tied to federal rules, in 21 CFR 1020:

  • Source-to-skin distance: at least 38 cm on stationary units and at least 30 cm on mobile units.
  • Beam-on signal: an audible signal at 5 minutes of irradiation time. It’s a prompt, not a dose limit.
  • Air kerma rate: generally no more than 88 mGy per minute, or 176 mGy per minute with high-level control active, at the measurement point.
  • Last-image hold: required on units made on or after June 10, 2006.

These working habits come from Sherer and Bushong:

  • Use pulsed mode at the lowest rate that gives a usable image, and keep beam-on time in short taps.
  • Use last-image hold to study a frame instead of exposing again.
  • Know what high-level control costs. Sherer says the exposure rate in this boost mode may significantly exceed the routine fluoroscopy rate.
  • Use magnification only when asked. On an image intensifier it raises mA and patient dose. On a digital system a zoom may not change exposure, so the physicist or manufacturer has to tell you.
  • Keep the curtain and Bucky slot cover in place.
  • Step back when you can, or stand behind the radiologist or in the booth. If your back could face the source, Sherer says a wraparound apron protects best.

On a C-arm:

  • Put the tube under the patient. With the tube on top, scatter and staff exposure rise.
  • Keep the image intensifier as close to the patient as you can.
  • For a cross-table shot, stand on the image intensifier side. Sherer says scatter on the tube side is typically 2 to 3 times higher.
  • Wear a 0.5 mm apron and a thyroid shield, and use the cable’s full length to gain distance.
Timeline over one second comparing continuous fluoroscopy with pulsed fluoroscopy at 30, 15 and 7.5 pulses per second
Fewer pulses per second means less beam-on time.

The 2027 outline splits the fluoroscopy positioning item into patient positioning and varying the beam angle, and groups cumulative air kerma, dose rate display and time documentation as dose monitoring equipment.

Where do you stand for a mobile exam?

Stand at least 2 m from the patient, the tube and the useful beam, and wear an apron. That is Sherer’s wording, and Bushong gives 2 m from the tube. Merrill’s writes 6 feet (2 m). Strictly, 6 feet is about 1.8 m. ARRT uses SI as its primary units, so learn 2 m first. We found no such distance in 21 CFR 1020. NCRP Report 102 says the operator shall stand at least 2 m (6 ft) from the patient, the x-ray tube and the useful beam, so this is an NCRP recommendation, not a federal rule.

Where you can, also stand at a right angle to the line between the beam and the patient, where scatter is lowest. Sherer says to deal with distance and shielding first.

Side view of a mobile x-ray unit with the radiographer in an apron at least 2 m from the patient, holding the exposure cord, at a right angle to the beam
At least 2 m, apron on, at a right angle to the beam-patient line.

Bontrager says only people needed for the exam should stay in the room, and anyone who stays gets protective apparel. Each mobile unit should have its own apron, as Sherer and Bushong both say.

How do you minimize patient exposure?

Good technique does the heavy lifting. Both outlines list these under the patient heading:

  • Exposure factors: kVp, mAs and AEC.
  • Beam restriction: collimate to the area of interest. Tighter than the receptor is allowed and often better.
  • Filtration, grids and the image receptor.
  • Patient considerations: positioning and clear communication prevent repeats. See the most common cause of repeat exposures. The 2027 list adds pregnancy and swaps “morbid obesity” for “body habitus”.
  • Dose documentation and the DAP meter.
  • Pediatric care: Image Wisely and Image Gently are named as resources in the 2027 outline.

A possibly pregnant patient. Ask before you expose, and ask the date of the last menstrual period. Tell the radiologist if the answer is yes or unsure. Sherer strongly recommends a pregnancy test before substantial pelvic exposure if there’s doubt. Then follow your facility’s protocol.

The two timing rules are worth knowing by name, and the sources don’t agree on them:

  • 10-day rule: schedule pelvic and lower abdominal exams in the first 10 days after menstruation starts. Ehrlich and Coakes says pregnancy is least likely then. Bontrager calls the rule obsolete, because canceling a needed exam can do more harm than the fetal dose. Merrill’s still uses it for bone densitometry. Sherer reports both NCRP advice to book nonurgent abdominal exams in the first few days after menses and an ACR position that they need not be postponed.
  • 28-day rule: none of those five books states it. The UK’s Royal College of Radiologists describes it this way: if a period is overdue and the patient can’t be sure she isn’t pregnant, consider postponing, usually after talking to the radiologist. The same 2013 guidance suggests limited use of the 10-day rule for high-dose exams such as barium enema and abdominal or pelvic CT. That is UK guidance, not a U.S. rule.

Bontrager adds a useful sorting rule: above the diaphragm or below the hips, scattered fetal dose is very low and the exam may proceed normally.

AAPM CARES quotes the American College of Obstetricians and Gynecologists: “With few exceptions, radiation exposure through radiography, computed tomography scan, or nuclear medicine imaging techniques is at a dose much lower than the exposure associated with fetal harm.”

Are gonadal and fetal shields still used?

Routine use is ending, but the answer depends on who you ask and where you work. Sherer (2018), Bushong (2017) and Ehrlich and Coakes (2021) all still teach shielding, and Ehrlich and Coakes calls it required in some cases. The statements below are newer or carry more authority, so go with them over the books:

  • April 2019, AAPM: routine patient gonadal and fetal shielding should be discontinued. Shields can obscure anatomy and interfere with AEC, and any benefit is minimal to nonexistent.
  • January 12, 2021, NCRP Statement 13: routine gonadal shielding isn’t justified in abdominal and pelvic radiography. NCRP’s companion document notes that some state rules still require it.
  • January 2021, ASRT: it’s appropriate to stop shielding for abdominal and pelvic imaging. ASRT still supports shields where placement is appropriate, such as a lap shield for an extremity exam.
  • Spring 2021, ARRT: institutions may adopt these recommendations to varying degrees, and programs should still teach shielding. ARRT’s exams may not refer specifically to gonadal and fetal shielding procedures, but they will cover patient radiation exposure and protection.
  • February 21, 2023, FDA: the gonad shielding recommendations in 21 CFR 1000.50 were removed.

Patient shielding isn’t listed under “Minimizing Patient Exposure” in either outline. ARRT also says it avoids content where resources conflict. So learn why routine shielding ended, know that the older technique exists, and follow state rules and facility protocol at work.

If a patient insists on a shield, AAPM CARES says it’s at the technologist’s discretion, as long as image quality isn’t compromised and overall dose isn’t increased.

What do dose-reduction scenarios look like?

Practice with scenarios like these. They are our examples, not ARRT items.

ScenarioCommonly taught action
The field is open wider than the anatomy of interestCollimate to the area of interest
A patient may be pregnant and has an abdomen series orderedStop and tell the radiologist. Follow protocol.
You’re close to the patient on a mobile examStep back to at least 2 m, at a right angle, with your apron on
A child is booked at an adult techniqueUse a pediatric technique and collimate tightly

What units does the exam use?

ARRT says SI units are the primary units on the exam. The 2027 outline cites NCRP Report 184 for them, where the 2022 outline cited Report 160.

QuantitySI unitOlder unitConversion
Absorbed doseGray (Gy)rad1 Gy = 100 rad
Dose equivalent and effective doseSievert (Sv)rem1 Sv = 100 rem
ExposureC/kgRoentgen (R)SI unit has no special name
Air kermaGray (Gy)Dose area product is air kerma times area

For x-rays, the NRC quality factor is 1, so 1 Gy of x-rays equals 1 Sv. For alpha particles it’s 20, and for neutrons of unknown energy it’s 10. Effective dose sums dose across tissues, weighted by each tissue’s sensitivity, and it’s used to compare risk across exams. Our formula toolkit has the dose conversions, which are part of a paid plan.

What else changes in this section on March 1, 2027?

Besides the question count and the items above:

  • The patient heading becomes “Minimizing Patient Exposure and Repeats”.
  • “Image receptors” becomes “digital detector”, with DQE and exposure latitude as examples.
  • Personnel monitoring now reads “recommendations and regulations”, citing NCRP Report 180 and 21 CFR.

The 2027 changes page covers the rest of the exam. For the digital side, including CR leaving the outline, read our Image Production guide.

How should you study Radiation Protection?

Learn the science first, then the rules. Read the Radiation Physics and Radiobiology chapter, then the Radiation Protection chapter. Then drill the Radiation Protection practice set, which has 317 questions today and from March 1, 2027. Every question has a written rationale. Infection control and contrast reactions sit in the sister guide, ARRT Patient Care.

The free tier is 3 chapters (Skull, Elbow and Shoulder, Urinary), 50 questions and a timed 200-question mock, with no card. Those are Procedures chapters, so no Safety chapter is free. You can still sign up free and use the mock to find your gaps. The two Safety chapters come with a paid plan: $39 a month, which renews until you cancel, $99 for a 90-day pass with no renewal, or $199 a year (about $16.58 a month), which renews yearly until you cancel. Paid plans open soon.

Frequently asked questions

How many Radiation Protection questions are on the ARRT exam?

There are 29 scored questions through February 28, 2027, and 25 from March 1, 2027. Radiation Protection sits inside the Safety section with Radiation Physics and Radiobiology, which stays at 21.

What is the occupational dose limit for the lens of the eye?

The NRC rule, 10 CFR 20.1201, sets 150 mSv (15 rem) a year. NCRP Commentary 26 (2016) recommended lowering NCRP's own limit to an absorbed dose of 50 mGy a year, which doesn't change the NRC rule. ARRT's 2027 outline names NCRP Report 180 for personnel monitoring, and we give no value from that report.

Does ALARA apply to patients on the ARRT exam?

ARRT's outlines list ALARA only under Personnel Protection and note that the concept is specific to personnel. They also say radiographers must apply protection principles to minimize bioeffects for both patients and personnel.

Do radiographers still use gonadal shields?

AAPM advises against routine gonadal and fetal shielding. NCRP says routine gonadal shielding isn't justified in abdominal and pelvic radiography, and FDA removed its recommendation in 2023. ASRT supports stopping it for abdominal and pelvic radiography. ARRT says institutions may vary, so follow your state rules and facility protocol.

What is the dose limit for a pregnant radiographer?

The NRC limit is 5 mSv (0.5 rem) to the embryo-fetus over the entire pregnancy, after the worker declares the pregnancy. Declaring is voluntary and must be in writing. Sherer and Bushong also give NCRP's monthly figure of 0.5 mSv, which is not in the NRC rule.

How far should you stand from the patient during a mobile exam?

Sherer says at least 2 m from the patient, the tube and the useful beam, and Bushong says at least 2 m from the tube, with an apron on. Merrill's writes it as 6 feet (2 m). Where you can, also stand at a right angle to the beam-patient line.

Sources

Found an error? Email [email protected] with the page and the source.

  1. ARRT Radiography Examination Content Specifications, 2022 and 2027: Question counts, outline topics, the ALARA note, SI units as primary, and the NCRP report each outline cites. Repo copies in docs/reference/arrt.
  2. 10 CFR Part 20, Standards for Protection Against Radiation: ALARA definition, sections 20.1201, 20.1208, 20.1301 and 20.1502, and units of dose. Repo copy in docs/reference/primary.
  3. 21 CFR Part 1020, Performance Standards for Ionizing Radiation Emitting Products: Fluoroscopy source-skin distance, air kerma rate limits, 5-minute signal and last-image hold. Repo copy in docs/reference/primary.
  4. NRC Issue Paper 2, Occupational Dose Limit for the Lens of the Eye: NRC and ICRP lens limits, and eye protection with side shields. Repo copy in docs/reference/primary.
  5. Dauer et al., overview of NCRP Commentary No. 26, Int J Radiat Biol, 2017: NCRP's recommended reduction from 150 mSv equivalent dose to 50 mGy absorbed dose. Abstract read. Commentary 26 itself not read.
  6. ARRT, Statement on Gonadal and Fetal Shielding, R.T. Update, Spring 2021: ARRT's four positions on shielding.
  7. AAPM Position Statement on Patient Gonadal and Fetal Shielding, PS 8-A: Recommendation to end routine shielding, with reasons.
  8. AAPM CARES, Patient Gonadal and Fetal Shielding FAQ: Pregnant and pediatric patients, and what to do if a patient insists. Repo copy in docs/reference/primary.
  9. NCRP, Implementation Guidance, companion to Statement No. 13, January 12, 2021: Routine gonadal shielding not justified in abdominopelvic radiography; some states still require it. Repo copy in docs/reference/primary. Statement 13 itself not read.
  10. ASRT Statement, January 12, 2021: Appropriate to discontinue shielding for abdominal and pelvic imaging.
  11. ASRT Update, January 15, 2021: Continued shielding where placement is appropriate, such as a lap shield for an extremity exam.
  12. FDA final rule, 88 FR 3638, January 20, 2023: Removed the gonad shielding recommendations in 21 CFR 1000.50, effective February 21, 2023.
  13. Mettler et al., summary of NCRP Report No. 184, Radiology, 2020: CT share of collective patient dose in 2016.
  14. Royal College of Radiologists (UK), Radiation and the early fetus, approved February 28, 2013: What the 28-day rule is, and limited use of the 10-day rule for high-dose exams. UK guidance, not a U.S. rule.
  15. NCRP Report No. 102, Medical X-Ray, Electron Beam and Gamma-Ray Protection for Energies Up to 50 MeV, 1989, Sections 3.3.4(p) and (q): At least 0.5 mm lead equivalent for aprons in the fluoroscopy room. A fluoroscopist's hand shall not be placed in the useful beam unless the beam is attenuated by the patient and a glove of at least 0.25 mm. Public copy read, cited without a link.
  16. Health Physics Society, Ask the Experts, question 14090: Minimum apron thickness is specified in state regulations, commonly 0.25 mm in some states and 0.5 mm in others.

Related posts

Patient Care

ARRT Patient Care: What the 33 Questions Test

ARRT Patient Care is 33 of 200 scored questions, 31 from March 2027: ID, vitals, infection control, contrast reactions, ethics and consent.

Study Strategy

How to Pass the ARRT Radiography Exam: A 60-Day Plan

A 60-day ARRT Radiography study plan: questions per section, what to study first, the score you need, and what changes on March 1, 2027.

Image Production

ARRT Image Production: What the 51 Questions Test

ARRT Image Production is 51 of 200 scored questions: exposure factors, grids, digital imaging and QC, plus what leaves on March 1, 2027.

Quick Answers on this topic

Direct sourced answers to questions readers most often follow up with.

Quick Answer

Two-Dosimeter Formula: 1.5 Waist + 0.04 Collar (NCRP 122)

NRC Regulatory Guide 8.40 credits NCRP 122 with the formula 1.5 x waist + 0.04 x collar. It estimates effective dose equivalent. ARRT lists no formula.

Quick Answer

Is a Thyroid Shield Required for Fluoroscopy?

No federal rule we read requires a thyroid shield in fluoroscopy. Your state or facility may. Sherer and Bontrager disagree on whether it is required.

Quick Answer

Most Common Cause of Repeat Exposures: Positioning Errors

Positioning leads rejected digital radiographs in published audits, at 49% to 79% where it is reported alone. Motion and exposure sit far behind.

Quick Answer

Lead Apron Attenuation: What the Numbers Really Say

Bushong lists 88% at 75 kVp and 75% at 100 kVp for a 0.5 mm Pb apron. A 2018 study measured about 97% at 100 kVp. No federal rule we read sets a minimum.

Turn the dose limits into reflex

Start free with 3 chapters, 50 questions and a timed 200-question mock, no card. The two Safety chapters, all 317 Radiation Protection questions and the full formula toolkit come with a paid plan, and paid plans open soon.

Report a bug

Tell us what's wrong. We'll take a look.